|
Preparation
of low-level QC serum
Start with 500ml of clear serum. Measure potassium
level. Calculate the final volume, to which 500 mg of this serum must be
diluted to adjust the potassium level to 3.5 mmol/L.
Initial volume 500 ml
Final volume (x) ml
Initial K+ value (e.g) 6.3 mmol/L
Final K+ 3.5 mmol/L
Therefore, 500 x 6.3 = 3.5 x (x)
x= 500 x 6.3/3.5 = 900 ml
15% of 900 ml = 135 ml
Extra
volume to be) = 400 ml
added (900-500
Ethanediol = 135 ml
Therefore,
distilled water
to be added (400-135) = 265 ml.
In actual practice, although potassium concentration
will be diluted to 3.5 mmol/L,
the levels of some other constituents will become too low. Therefore the
levels of these need to be raised to some extent so that these will be
maintained at low level but not too low. Out of the 265 ml distilled water,
25 ml may be used for dissolving desired quantities of such of those
constituents that need to be added to the serum to raise their amount to the
desired low levels.
Adjustment
of analytes
Glucose, urea, creatinine,
sodium (NaCl) and potassium (KCl):
-
Dissolve each of these in 5 ml distilled water and add
to the main bulk of serum.
Calcium :- Use dried CaCO3.
Dissolve in 5 ml in 1N HCL and add to the serum.
After adding all the analytes
mix the contents well and centrifuge the serum once again for 10 minutes at
3500 rpm. Collect the supernatant serum in a sterile one-litre
flask. Mix the contents well and dispense the serum into 5 ml sterile
penicillin vials, seal the vials and store at –200C. Laboratories
can decide on the volume of aliquots depending on their requirements.
Construction of Levey
Jennings Chart
On each day when analyses are performed a fresh sample
is thawed, thoroughly mixed and analysed. (Remember:
Ethanediol-treated serum may not freeze
completely at –200C; however, the constituents are quite stable).
The QC serum is analysed for a period of 20 days or
so. [Important to note: Analysis should not be carried out by only one
person; all staff should participate in this exercise to determine the true
unavoidable error in the laboratory]. From these data, mean and SD are
calculated. Levey Jennings chart is then constructed with x
+ 2SD as warning limits and x + 3 SD as control limits.
Calculate the %CV for each analyte
to ascertain whether this is within the acceptable limit (Ideal = <
5%. Must be definitely < 8%). If % CV is found to be high, this
will indicate that between-day laboratory precision (variation) is high and
the data cannot be used to construct a Levey
Jennings chart. It is then essential to identify the causes for this, correct
these and then repeat the whole exercise and confirm that the %CV is well
within the acceptable limit.
Table 1 shows precision data obtained
by a WHO trainee from a developing country for routine biochemical analytes through analysis of ethanediol-treated
QC serum for 20 days, while undergoing training at the author’s laboratory
recently. Table 1
|
Analyte
|
Autoanlyser Hitachi 912
|
Manual Method
|
|
Mean
|
%CV
|
Mean
|
%CV
|
|
Glucose
|
167
|
1.2
|
166
|
4.2
|
|
Total protein
|
9.8
|
1.2
|
9.6
|
3.3
|
|
Albumin
|
3.5
|
1.4
|
3.6
|
4.2
|
|
Urea
|
50
|
1.7
|
55
|
3.8
|
|
Calcium
|
10.3
|
1.9
|
10.5
|
4.0
|
|
Cholesterol
|
185
|
1.4
|
180
|
4.5
|
|
Creatinine
|
1.7
|
0.7
|
1.8
|
3.0
|
Interpretation of QC data
A. According to WHO(6)an analytical system is ‘out of control’ if one of the four
criteria is met. That is:
a value lies entirely outside the control
limits
seven consecutive values show a rising
tendency
seven consecutive values show a falling
tendency
seven consecutive values lie on the same side
of the mean
If one of these situations arises, the patients’ results
must be discarded, the cause of the error sought and removed, and then the
batch repeated with a QC serum.
B. Use of two different levels of QC simultaneously in
every batch of analysis provided valuable information on the type of errors –
whether these are random (precision) errors or systematic (accuracy) errors.
Westgard’s rules for interpreting QC data
obtained using a single-level QC as well as two different-level QCs are
schematically presented in Fig.1.(7)

Warning Rule
12S One observation > x + 2 SD
Rejection Rules (used if warning rule is exceeded; run rejected
if any of the following rules are violated) R= Random error; S =
Systematic error
R 13S
One observation > x + 3 SD
S 22S
Two observations > same limit, that is x + 2 SD or x - 2SD (same control-
two consecutive runs, or two different controls – same run)
R R4S
Difference between two observations within run > 4SD (two different
controls – one > x + 2SD and the other > x – 2SD)
S 41S
Four consecutive observations>same limit, that is x + 1SD or x - 1SD (same
control four consecutive runs )
S 10X
Ten consecutive observations on same side of mean (same control, ten
consecutive runs, or two different controls, five consecutive runs)
Remedial action
A well-run internal QC system makes possible immediate
intervention in the release of patients’ results. In the event of a control
system alert, it is advisable to proceed through the following steps in that
order.
Decision:Immediate
decision whether action is necessary.
Investigation:Check to locate the error.
Repair:Action
to eliminate the error.
If the decision is taken that the method is out of
control, the first action is to withhold patients’ results in that batch.
Then the analyst should start by checking for the simplest and most frequent
faults, and then continue as necessary in a logical order depending on the
method and equipment involved.
It is good practice to start by excluding gross errors
such as mix-up of control materials, reagents or pipettes, misuse of
measuring instruments [wrong filter or aged lamp in the photometer], or
failure to follow instructions for a step in the method.
The results obtained on a control specimen may be in
error for several reasons, including deterioration due to age, incorrect
storage or contamination, wrong identification and mistake in preparation or
constitution.
Further action will depend on whether the alert is due
to a change in accuracy or in precision. If accuracy has deteriorated,
attention should be focused on the possibility of systematic sources of error
such as incorrect reaction temperature, calibration errors and faulty devices.
If precision has deteriorated, steps in the analytical procedure should be
checked, for instance, deproteinization,
composition of reagents and reaction mixtures, measuring systems [photometer,
flame photometer, etc].
To easily differentiate between systematic and random
errors, laboratories are encouraged to construct Youden
charts. These are constructed by having x + 2SD & x + 3SD
of one level QC in the x axis and that of another level QC in the y axis. If
an analyst analyses both levels of QC and plots the data in the Youden chart, a single plot will be obtained. If this
falls within the inner square, it means that the value
obtained for both QC are well within the acceptable limit, i.e. x +
2SD. On the other hand, if a plot appears outside this limit, it could mean
that the data are outside the acceptable limit and the error could be either systematic or random. Fig.2 shows sample data
plotted in Youden chart. Figure 2.

It
can be inferred that :-
1. Plots E, F, G, K, M, N, O & R are
acceptable
2. Plot A values for both QC are > +2SD
Plot C values for both QC are > -2SD
3. These data indicate 22S error, i.e.
systematic error
4. Plot H Values for QC1 (
x axis) is > - 2SD
while value for QC2 (y axis) is > + 2SD
This indicates R4s error i.e. random error
Prevention of systematic errors
To prevent or minimize systematic errors, the laboratory
should adhere to the following points:
Use of proper calibration technique. Use of
pure chemicals, precision balance, quality distilled
water. Proper preservative and storage
Regular checking of photometric filter, bulb,
tubing, etc.
Use of recommended analytical methods
Calibration of pipettes at regular intervals
Instrument calibration – photometric check
Use of calibrated cuvette
Regular preventive maintenance of equipment –
daily, weekly and monthly.
While it is easy to identify systematic errors, it is
quite difficult to pinpoint random errors. In order to minimize this
possibility, it is important to educate the staff on various aspects that
could lead to random errors.
External quality assessment
Participation in External Quality Assessment Schemes is
important for interlaboratory comparison and the
maintenance of long-term accuracy and precision of the analytical systems in
the laboratory. This gives the laboratory an opportunity to have an appraisal
of the methods employed and switch over to better methods.
References
1. Nordtest Technical
Report from the technical group for quality assurance – Report #187 (1992).
2. NABL (National Accreditation Board for Testing
and Calibration Laboratories, India)
criteria for laboratory accreditation – NABL #101 – 2nd Edition (1994).
3. NABL Specific guidelines for accreditation of
clinical laboratories and checklist for assessors (1998).
4. Kanagasabapathy
AS, Swaminathan S. and Selvakumar R. Quality Control in Clinical Biochemistry
Indian Journal of Clinical Biochemistry (1996); 11: 17-25.
5. Browning DM, Hill PG and Olazabal
DAV. Preparation of Stabilized liquid quality control serum to be used in
clinical chemistry –WHO document LAB/86.4
6. Stamm D. Guidelines
for a basic programme for internal quality control of quantitative analysis
in clinical chemistry – WHO document LAB/81.3.
7. Westgard, Barry
PL and Hunt MR. A
multi-rule shewart chart for quality control in
clinical chemistry. Clin. Chem. (1981); 27:493-501.

|