|
To confirm a clinical suspicion of intracompartmental
syndrome (see 5.8.2) the pressure inside the particular compartment
should be measured directly.
The
threshold pressure required to initiate the flow of liquid into the fascial compartment is a measure of the tissue pressure
inside that compartment. With full sterile precautions and after infiltrating
local anaesthetic, a 21 or 22 gauge cannula, approximately 3-4 cm long, is inserted into the
compartment through or around an introducing 20 or 21 gauge needle. The cannula is connected
through narrow pressure tubing to a syringe or low speed infusion pump.
Through a three-way tap, the system is connected, through a side arm to a blood pressure transducer or saline or mercury manometer (Fig
47). The system is filled with sterile isotonic saline. If a syringe-type
infusion pump and arterial blood pressure transducer with monitor is used,
the pressure can be measured continuously at a very slow rate of infusion (eg 0.7 ml/day). If a saline or mercury manometer is used,
a much higher rate of infusion is required to initiate flow into the
compartment. These systems are not suitable for continuous intracompartmental pressure monitoring.
|

|

|
| Figure 47 Infusion pump,
saline manometer system in use for measuring the tissue pressure inside the
anterior tibial compartment (Copyright DA Warrell)
|
Figure 48Stryker
pressure monitor in use for measurement of intracompartmental
pressure (Copyright DA Warrell)
|
Alternatively, the simple but expensive Stryker pressure monitor can be used (Fig 48).
Whatever system is employed, the zero point in the
pressure measuring device must be aligned to the level at which the cannula enters the fascial
compartment.
|