Guidelines on Standard Operating Procedures for CLINICAL CHEMISTRY

Urea – Diacetyl monoxime method

 

*     Introduction

Urea contributes most of the body’s non-protein nitrogen, accounting for about 45% of the total. It is the major end-product of protein catabolism in humans. It is synthesized in the liver, released into blood circulation and excreted by the kidneys. Measurement of urea in blood is a useful indicator of renal and hepatic integrity.

*     Principle of the method

Urea reacts directly with diacetyl monoxime under strong acidic conditions to give a yellow condensation product. The reaction is intensified by the presence of ferric ions and thiosemicarbazide. The intense red colour formed is measured at 540nm/ yellow green filter.

*     Specimen types, collection and storage

Serum is the specimen of choice. Store samples for no longer than 8 hours at room temperature (25-350C) and 7 days at 2-80C. For a longer duration, store in the freezer. If the samples show evidence of bacterial contamination, do not use these for urea estimation. Plasma could also be used for urea estimation.

*     Reagents

All chemicals must be Analar grade.

*     Stock acid reagent

Dissolve 1.0g of ferric chloride hexahydrate in 30 ml of distilled water. Add 20 ml orthophosphoric acid and mix. Store in a brown bottle at room temperature (25-350C).Stable for 6 months.

*     Mixed acid reagent

Add slowly 100 ml of Conc. H2S04 to 400 ml distilled water taken in a 1-litre flat-bottom conical flask kept in an icecold waterbath. Mix well and add 0.3ml of stock acid reagent. Mix and store in a brown bottle at room temperature (25-350C). Stable for 6 months.

*     Stock colour reagent – A

Dissolve 2g diacetyl monoxime in distilled water and make the volume up to 100 ml in a volumetric flask. Store in a brown bottle at room temperature (25-350C). Stable for 6 months.

*     Stock colour reagent - B

Dissolve 0.5 g thiosemicarbazide in distilled water and make up to 100 ml in a volumetric flask. Store in a brown bottle at room temperature (25-350C). Stable for 6 months.

*     Mixed colour reagent

Mix 35 ml of stock colour reagent A with 35 ml of stock colour reagent B and make up to 500 ml with distilled water. Store in a brown bottle at room temperature (25-350C). Stable for 6 months.

*     Stock urea standard

Weigh 1.0g of analytical-grade urea and dissolve in 100ml of benzoic acid (1g/dl). Use a 100ml of volumetric flask for preparing this. Store at room temperature (25-350C). Stable for 6 months.

*     Working standard 50mg/dl

Dilute 5.0ml of stock urea standard to 100 ml with benzoic acid. Store at room temperature (25-350C). Stable for 6 months.

*     Equipment, glassware and other accessories

Refer to Section A (2), Introduction to SOP

*     Procedure

The protocol of the procedure is described below.

*     Dilution of Standards (S1-S3), Test & QC

Pipette the following into appropriately labelled 13 x 100 mm tubes

 

S1

S2

S3

Test

QC

Distilled Water (ml)

1.9

1.8

1.7

1.9

1.9

50 mg/dl Urea (ml)

0.1

0.2

0.3

-

-

Test sample /QC (ml)

-

-

-

0.1

0.1

Mix Well

*     Colour Development

The colour reagent is prepared fresh at the time of analysis by mixing distilled water, mixed acid reagent and mixed colour reagent in the ratio 1:1:1.

Pipette the following into another set of appropriately labelled 18 x 150 mm tubes.

 

Blank

S1

S2

S3

Test

QC

Colour reagent (ml)

3.1

3.0

3.0

3.0

3.0

3.0

Respective diluted standard ml)

-

0.1

0.1

0.1

-

-

Diluted test /QC (ml)

-

-

-

-

0.1

0.1

Mix all tubes well. Keep them in a boiling waterbath for 15 minutes. Remove from waterbath and cool the tubes for 5 minutes. Set the spectrophometer/filter photometer to zero with blank at 540nm/yellow green filter and measure the absorbance of the other tubes.

*     Calculation and calibration graph

Concentration of standards:

S1 = 50 mg/dl
S2 = 100 mg/dl
S3 = 150 mg/dl

Plot the absorbance values of standards against their respective concentrations. The measurable range with this graph is from 10 to 150 mg/dl. A calibration graph should be constructed whenever a new set of reagents is prepared. Plot absorbance values of test/QC on the calibration graph and read off the concentrations.

Once linearity is proved, it will be enough if S3 is set up every time that patients’ samples are analysed and the results calculated using the formula:

Absorbance of test
Urea in test sample = ----------------------------- x 150 mg/dl
Absorbance of Standard

*     Analytical reliabilities

Refer to pages 7-9 of section 1 (General Introduction) for the use of internal QC and interpretation of daily QC data (for releasing patients’ results).

Since urea is one of the most common analytes measured in a laboratory after glucose, it is recommended that an internal QC (normal QC pool) be included with every batch of samples analysed in the day, irrespective of the number of samples in a batch. Further, even when a single sample is analysed as an "emergency" sample at any time of the day or night, it is essential to include an internal QC. From the QC results obtained for the day, mean, standard deviation and %CV can be calculated to ensure that within-day precision is well within the acceptable limit, i.e. 4%.

The mean value of internal QC for the day can be pooled with the preceding 10 or 20 mean values obtained in the previous days and between–day precision can be calculated and expressed as % CV. Ensure that this is well within the acceptable limit, i.e. 8%.

At least once a day analyse another QC serum from either a low QC or high QC pool.

"Assayed" QC sera with stated values (ranges) are available from several commercial sources, viz. Boehringer Mannheim, BioRad & Randox.

If a laboratory uses QC sera from a commercial source, it is important that the company certifies that their QC materials are traceable to international reference materials.

*     Hazardous materials

Most of the chemicals used in this method are acids. Care should therefore be taken to avoid mouth pipetting and contact with skin.

*     Reference range and clinical interpretation

Serum/ Plasma Urea ……..15 – 40 mg/dl

Elevated serum urea levels may be due to pre-renal, renal or post-renal etiology. Pre-renal causes could be cardiac related or due to increased protein catabolism, and dehydration. Renal causes include glomerulonephritis, chronic nephritis, nephrotic syndrome and other kidney disease. Post-renal causes include obstruction of the urinary tract.

Decreased serum urea levels could be due to pregnancy, intravenous infusion, low antidiuretic hormone secretion, low protein intake, severe liver diseases, inborn errors of urea cycle and SIADH (Syndrome of inappropriate ADH secretion).

*     Limitations

Specimens with gross icterus cannot be assayed as it will cause falsely elevated urea values. Do not report results from specimen with suspected interference. Inform the requesting physician of the problem.

*     References

 

1.       Wybenga, D.R., Di Glorgio, J.& Pileggi, V.J. (1971). Clinical Chem., 17, 891-895.

2.       Seaton, B & Ali. A (1984) Med. Lab. Sciences 41, 327 – 336.

 

 

||| | ||