|
Principle of the method
Glucose present in the CSF is oxidized by the enzyme
glucose oxidase (GOD) to gluconic
acid with the liberation of hydrogen peroxide, which is converted into water
and oxygen by the enzyme peroxidase (POD). 4 aminophenazone, an oxygen acceptor, takes up the oxygen
and together with phenol forms a pink coloured chromogen which can be measured at 515nm.
GOD
Glucose 4 Gluconic acid + H2O2
-------------------> 4 Gluconic acid + H2O2
POD
H202--------------> 4
H20 + [0]
[0] + 4 aminophenazone + phenol
------> chromogen (pink)
Reagents
All chemicals must be Analar
grade
Phosphate
buffer : 100 mmol/L. pH 7.0
To 800 ml of distilled water add the following in the
order:
Disodium hydrogen phosphate dihydrate
[Na2HPO4 2H2O]…..12.95 g;
Anhydrous potassium dihydrogen phosphate [KH2PO4]…..4.95
g;
Sodium azide [NaN3]…..0.5 g
Add one by one, dissolve and finally make up to one litre with distilled water. Stable for 3-4 months, at 2-80C.
Check final pH with a pH meter.
Colour reagent
To 100ml of the above phosphate buffer add the following
in the order and then mix to dissolve :
4
amino phenazone 16 mg
GOD [Sigma G 7016] 1800 units
POD [Sigma P 8250 ] 100 units
Phenol 105 mg
Tween 20 [Sigma P 1359] 50m l
Reconstitute the GOD and POD powder with phosphate
buffer. Dispense separately into vials so that each vials represents the
requisite number of units. Store the vials frozen.
Stable for 2 weeks at 2-80C. Store in a brown bottle.
Benzoic
acid 1g/l.
Dissolve 1.0g of benzoic acid in water and make up to
one litre with water. This solution is stable
indefinitely at room temperature.
Stock
Glucose solution, 1 g/l.
Before weighing, dry the glucose at 60-800C
for 4 hours. Allow to cool in a dessicator.
Dissolve 1g of glucose in benzoic acid solution and make up to 100 ml in a
volumetric flask. Stable for six months at room temperature (25-350C).
Do not freeze the standard
Working
glucose standard 100 mg/dl.
Dilute 10 ml of stock glucose (use either a volumetric
pipette or a burette) to 100 ml with benzoic acid in a 100 ml volumetric
flask. Mix well. Stable for 6 months at room temperature (25-350C).
Equipment, glassware and other accessories
Refer to Section A (2),Introduction to SOP.
Procedure
The protocol of the procedure is described below.
Dilution
of standards (S1-S5), test & QC
Pipette the following into appropriately labelled 13 x 100 mm tubes
|
|
S1
|
S2
|
S3
|
S4
|
S5
|
Test
|
QC
|
|
Distilled Water (ml)
|
1.9
|
1.8
|
1.7
|
1.6
|
1.5
|
1.9
|
1.9
|
|
100 mg/dl glucose (ml)
|
0.1
|
0.2
|
0.3
|
0.4
|
0.5
|
-
|
-
|
|
Test sample /QC (ml)
|
-
|
-
|
-
|
-
|
-
|
0.1
|
0.1
|
Mix well
Colour development
Pipette the following into another set of appropriately labelled tubes.
|
|
Blank
|
S1
|
S2
|
S3
|
S4
|
S5
|
Test
|
QC
|
|
Colour reagent (ml)
|
1.2
|
1.2
|
1.2
|
1.2
|
1.2
|
1.2
|
1.2
|
1.2
|
|
Distilled water (ml)
|
0.1
|
0.1
|
-
|
-
|
-
|
-
|
-
|
-
|
|
Diluted Standards (ml)
|
-
|
-
|
0.1
|
0.1
|
0.1
|
0.1
|
-
|
-
|
|
Diluted Test sample/QC (ml)
|
-
|
-
|
-
|
-
|
-
|
-
|
0.1
|
0.1
|
Mix all tubes well. Incubate at 370C in a
waterbath for 15 minutes. Remove from waterbath and cool to room temperature. Set the
spectrophotometer/ filter photometer to zero using blank at 510 nm/ green
filter and measure the absorbance of standards, test and QC. This protocol is
designed for spectrophotometers / filter photometers that require a minimum
volume of reaction mixture in the cuvette of one
ml. or less. Since economical use of reagents is possible with this protocol,
the cost per test can be kept to the minimum. However, if a laboratory employs
a photometer requiring a large volume of the reaction mixture for
measurement, viz. 5 ml, it is advised that the volume of all reagents
mentioned under Tabulation "(b) Colour
development", be increased proportionately.
Calculation and calibration graph
Since the protocol for standard tube S1 and test is
identical, the standard S1 will represent a concentration of 100 mg/dl. The
glucose concentrations represented by other standard tubes are S2 =200; S3 =
300; S4 =400 and S5 = 500 mg/dl.
Plot the absorbance values of standards against their
respective concentrations. The measurable range with this graph is from 10 to
500 mg/dl.
Plot absorbance values of test/QC on the calibration
graph and read off the concentrations.
Once linearity is proved, it is not necessary to prepare
the standard graph every time that patients’ samples
are analysed. It will be adequate if standard S2 is
taken every time and patients’ results are calculated using the formula :
Test
absorbance
--------------------------- x 200 ………… mg/dl
Standard absorbance

Analytical reliabilities
Refer to pages 7-9 of section 1 (General
Introduction) on the use of internal QC and interpretation of daily QC
data (for releasing patients’ results).
Since CSF analysis is carried out infrequently in
intermediate laboratories, one QC for glucose should be included as and when
CSF glucose is analysed. Hence it will not be
possible to analyse several QC samples and calculate
within-day precision. However, even if only a single QC sample is analysed in a day, this value can be pooled with the
preceding 10 or 20 values obtained in the previous days and between-day
precision can be calculated and expressed as % CV. Ensure that this
is well within the acceptable limit, i.e, 8%.
At least once a day analyse
another QC serum from either a low QC or high QC pool.
"Assayed" QC sera with stated values (ranges)
are available from several commercial sources, viz. Boehringer
Mannheim, BioRad & Randox.
If a laboratory uses QC sera from a commercial source,
it is important that the company certifies that their QC materials are
traceable to international reference materials.
Hazardous materials
This procedure uses sodium azide and phenol, which are poisonous and caustic. Do not
swallow, and avoid contact with skin and mucous membranes
Reference range and clinical interpretation
Concentrations of analytes in
the CSF should always be compared with those in plasma. Normal CSF glucose is
about 60% of the plasma value.
Normal range for CSF glucose 50-80 mg/dl
Decreased CSF glucose levels are observed in
tuberculosis, benign lymphocytic chronic meningitis
and in hypoglycemia. Increased levels are observed in encephalitis,
poliomyelitis and in cerebral abscess.
Limitations
Grossly bloody CSF may give spuriously elevated values
for glucose. Undue delay in analysis may give low values. The report to the
requesting physician should include the appearance of the CSF before and
after centrifugation.
References
1. Trinder, P. (I 969). Annals of Clin.
Biochem. 6: 24 - 27.
2. Barham D and Trinder P. (1972).
Analyst 97: 142-145.
|