|
Sterilization is
defined as the destruction or removal (by filtration) of all microorganisms
and their spores, whereas disinfection is the destruction of many
microorganisms but not usually the bacterial spores. Sterilization is usually
achieved with the help of heat whereas chemical agents are employed to effect
disinfection.
Sterilizationand disinfection are part of the daily routine of microbiological
laboratories and constitute a vital activity which ensures that cultures,
containers, media and equipment are treated in such a way that only the
inoculated organisms will grow while all others will be eliminated.
Sterilization
by heat
This can be achieved by autoclaving, by exposing
articles to dry heat in hot air ovens or boiling.
Autoclaving
Autoclaves can sterilize anything that can withstand a
temperature of 121oC for 30 minutes. A pressure cooker used in
homes for cooking purposes can also be used as a makeshift autoclave.
The containers having clinical material are subjected to
heat treatment in the autoclave after which these are emptied and washed and
put back into service.
Only autoclaves designed for laboratory work and capable
of dealing with a mixed load should be used. Porous load and bottle fluid
sterilizers are rarely satisfactory for laboratory work. There are two
varieties of laboratory autoclaves:
Pressure cooker type.
Gravity displacement models with automatic air
and condensate discharge.
Pressure-cooker type laboratory
autoclaves
The most common type is a device for boiling water under
pressure. It has a vertical metal chamber with a strong metal lid which can
be fastened down and sealed with a rubber gasket. An air and steam discharge
tap, pressure gauge and a safety valve are fitted in the lid. Water in the
bottom of the autoclave is heated by external gas burners, an electric
immersion heater or a steam coil.
Operating
instructions
Ensure that there is sufficient water inside
the chamber.
Load the autoclave and fasten the lid keeping the
discharge tap open.
Adjust the safety valve to the required
temperature and turn the heat on.
Allow the mixture of air and steam to pass out
freely till all air has been discharged.
Close the air discharge tap and let the steam
pressure rise within the chamber till it attains a temperature of 121oC (1.5 kg/cm2).
Hold on the pressure for 15 minutes.
Turn off the heat and let the autoclave cool.
Slowly open the air and steam discharge taps
after the pressure gauge has reached zero.
Allow the material to cool before these are
handled (usually agar bottles take hours before these become safe to handle).
Autoclave
with air discharge by gravity displacement
These are usually rectangular in shape and arranged
horizontally. These autoclaves have a jacket around the chamber.
Operating
instructions
Bring the jacket of the autoclave to operating
temperature.
Load the chamber, close the door and open the
steam valve so that steam can freely enter the top of the chamber. Air and
condensate shall automatically flow out through the drain at the bottom.
When the drain thermometer reaches the
required temperature, allow further period for the load to reach that
temperature (this has to be determined initially and periodically for each
autoclave).
Continue the autoclave cycle for the holding
period.
Close the steam valve and let the autoclave
cool till a temperature of 80oC is reached.
Gradually and softly open the autoclave
enabling the steam to escape and allow the load to cool further.
Hot air
oven
A hot air oven is electrically operated and should be
equipped with a fan to ensure uniform temperature inside. The required
temperature for sterilization is generally 160oC for one hour.
Operating
instructions
Arrange the material to be sterilized loosely
and evenly on the racks of the oven allowing free circulation of air and
thereby even heating of the load.
Do not pack the load tightly since air is a
poor conductor of heat.
Switch on the power supply and control the
temperature of the oven by adjusting the thermostat.
Note the time when the desired temperature is
reached (heating-up time).
Hold the load in the oven at this temperature
for a definite period of time (holding period). This is usually 60 minutes at
160oC.
Do not overheat since it would char the cotton
plugs and paper wrappings.
Autoclaves and hot air ovens can be used for
disinfection of infectious waste before it is discarded. In addition, waste
can be disposed of by boiling in detergent or by burial.
Boiling
in detergent
In the absence of an autoclave, most specimen containers
can be boiled in water having detergents to decontaminate. This process kills
the vegetative bacteria but fails to destroy the spores and certain viruses.
The easiest way to get best results is to add washing powder or sodium
carbonate crystals, 60 grams to one litre of water
in a big container and boil specimen containers in it for a minimum of 30
minutes.
Disinfection
Disinfection can be undertaken either chemically or by
boiling. Boiling is an effective method to disinfect equipment e.g. needles
and syringes, if autoclaving facilities are not available. Equipment which
has already been cleaned should be boiled for 20 minutes. Chemical
disinfection is used for heat-sensitive equipment that is damaged at high
temperatures. Commonly-used chemical disinfectants include chlorine releasing
compounds; ethyl and isopropyl alcohol, quaternary ammonium compounds and gluteraldehyde.
The synopsis of a few commonly-used disinfectants is
given in Table 1.
Preferred methods of sterilization for common articles
are given in Table 2.
Decontamination of some of the commonly reusable
equipment has been briefly presented in Table 3.
Table 1: Disinfectants and their
mode of application*
|
Target
|
Disinfectant
|
Strength
to use (disinfectant/
material V/V)
|
Application
|
Time
of exposure
|
|
Skin
|
Ethanol
Iodine
Povidone iodine
Quaternary ammonium comp
|
70%
1%
1%
|
Direct
Direct
Direct
Direct
|
2 minutes
2 minutes
2 minutes
2 minutes
|
|
Blood
|
Cresol
(pH 9)
Ca hypochlorite
|
5%
1%
|
2:1
2:1
|
6 hours
6 hours
|
|
Urine
|
Cresol
(pH 9)
|
5%
|
1:1
|
4 hours
|
|
Sputum
|
Cresol
(pH 9)
|
5%
|
1:1
|
4 hours
|
|
Faeces
|
Cresol
(pH 9)
Hypochlorite
(Na/Ca)
Ca hydroxide
|
5%
1%
20%
|
2:1
3:1
2:1
|
6 hours
6 hours
6 hours
|
|
Work benches
|
Lysol
Cresol
Hypochlorite
Chloramine-T
|
5%
1%
5%
|
Direct
Direct
Direct
Direct
|
4 hours
4 hours
4 hours
4 hours
|
|
Glassware
|
Hypochlorite
|
1%
|
Direct
|
4 hours
|
|
Lab instruments
|
Hypochlorite
Isopropanol
|
0.1%
70%
|
Direct
Direct
|
4 hours
4 hours
|
* Based upon: Basics of quality assurance: WHO/EMRO,
1992, page 162
Table 2: Preferred methods of sterilization for common-use articles
|
Autoclaving
Animal
cages
Sugar tubes
Lab. coats
Cotton
Filters
Instruments
Culture media
|
Hot
air oven
Glass ware
Beakers
Flasks
Petridish
Pipette
Slides
Glass syringes
Test tubes
Powders
|
|
Rubber
Gloves,
stopper, tubing
|
Wood
Tongue
depressor, applicator
|
|
Glass
Slides,
syringes, test tubes
Enamel metal trays
Wire baskets
|
|
Table 3: Disinfection of specific equipment
|
Container/material
|
Method
of choice for decontamination
|
Alternative
method of decontamination
|
|
Reusable stool container
|
Autoclaving 121oC for 30 minutes
|
Fill the jar having stool with 5% solution of phenol and keep for
24hours
|
|
|
Empty into lavatory*
|
Empty into lavatory*
|
|
Reusable containers of CSF, pus, sputum
|
Autoclaving
|
Boiling in detergent
|
|
Urine bottles (after emptying in lavatory*)
|
Autoclaving
|
Fill with 2% phenol or 1% bleach, leave for 4 hours, clean with
detergent
|
|
Blood containers
|
Autoclaving
|
Soak overnight in strong disinfectant(5% cresol; 1% Ca
hypochlorite, 1:2 V/V)
|
|
Glass microscope slides**
|
Autoclaving
|
Soak overnight in 5% phenol
|
* If the lavatory is connected to a septic tank,
phenol or other antiseptics should not be put into the lavatory.
** Glass microscope slides which have been used for
the diagnosis of tuberculosis should be discarded after keeping them soaked
in detergent overnight.
Biohazard waste management
Waste is defined as any solid, liquid or gaseous
material that is no longer used and will either be recycled, disposed of or
stored in anticipation of treatment and/or disposal.
Storage
Prior to disposal, all biohazardous
waste should be maintained and stored separately from the general waste
stream and from other hazardous wastes. The containers used to store biohazardous waste should be leak-proof, clearly labelled with a red or orange universal biohazard symbol
and sealed tightly when transported. In certain cases, it may be necessary to
double-bag the waste to prevent leakage. Any biohazardous
sharps, such as infectious needles and scalpels, must be placed in containers
that are puncture-resistant, leak-proof on all sides and the bottom, and
close-able. These containers can then be placed in a standard biohazard bag.
Disposal options
There are three main disposal options:
render the waste
noninfectious by autoclaving and dispose it in the general waste stream. If
autoclaving is not possible, decontaminate with chemical disinfectants or by
boiling for 20 minutes before disposal.
on-site incineration,
if possible.
transportation of
locally-generated waste to a distant appropriate facility.
Incineration is the preferred disposal option. Not only
does this method render the waste noninfectious but it also changes the form
and shape of the waste. Sterilization is an effective method for
decontaminating waste, but it does not alter the appearance of the waste.
Steam sterilization in an autoclave at a temperature of 121oC for at least 15
minutes destroys all forms of microbial life, including high numbers of
bacterial spores. This type of complete sterilization can also be accomplished
using dry heat which requires a temperature of 160-170oC for 2-4 hours.
However, it must be ensured that heat comes in contact with the material to
be rendered sterile. Therefore, bottles containing liquid material should
have loosened caps or cotton plug caps to allow for steam and heat exchange
within the bottle. Biohazard bags containing waste should be tied loosely.
Once sterilized, biohazardous waste should be
sealed in appropriate containers, labelled as
disinfected waste and disposed of in an approved facility.
Biological waste should be clearly labelled
prior to disposal and complete records should be maintained.
Burial
It is not a decontaminating process per se.
However, it does prevent the infectious material from becoming a reservoir of
infection if properly buried. It requires digging a pit of almost 5 meters
depth and 2 meters width and having a tightly fitted heavy lid on top.
Disposable containers with clinical material are thrown daily into it and the
lid is replaced immediately after throwing the specimens. Once a week, the
refuse is covered with a layer of quicklime. If quicklime is not available,
the refuse is covered with almost 10 cm thick layer of dried leaves once a
week.
Further
reading
1. El-Nageh MM et al. Basics of Quality Assurance for
Intermediate and Peripheral Laboratories. WHO Regional Publication, Eastern
Mediterranean Series No 2, 156-166, 1992.
|